M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
14.4
Technical specications and illustrations are without
obligation, subject to modications. 04.04/06.06
Technical Data
Gas Conditioning Unit series CSS
®
version CSS-M version CSS-M/W
Part No. for 230V 50Hz version 01G2000 01G2500
Part No. for 115V 60Hz version 01G2000a 01G2500a
Sample outlet dew point range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
Dew point stability at constant conditions < ± 0,1 °C**
Sample inlet temperature **max. 80 °C
Sample inlet dew point **max.+50 °C
Gas flow rate / flow alarm **max. 70 Nl/hr / no **max. 70 Nl/hr / optional part no. 01G9805
Ambient temperature **+5 °C up to +45 °C
Storage temperature -25 °C up to +65 °C
Pressure 0,7 bar up to 1,4 bar abs.
Total cooling capacity ** max. 40 kJ/hr
Number of gas inlets 1
Number of gas outlets 1
Condensate outlet 1
Medium connections tube connection 4/6 mm
Material of sample contacting parts glass, PVDF, Novoprene
®
,FPM
Ready for operation approx. 10 min.
Power supply 230V 50Hz or 115V 60Hz
Power consumption max. 240VA
Fuse protection 4A t, 5x 20 mm
Electrical connection Cold appliance plug with 2 m of cable mains cable 3x1mm
2
, signal cable 3x0,5mm
2
, 3m length
Status alarm, 1change over contact, poten-
tial free
no contact rating 250VAC, 2A, summary alarm for gas flow, li-
quid and cooler temperature
Case protection IP20 (DIN 40050. IEC 529)
Sheet steel housing, execution portable, colour grey, RAL 7035 wall mounting, colour grey RAL 7035
Equipment dimensions 177 mm x 220 mm x 300 mm (H x W x D) 177 mm x 263 mm x 372,5 mm (H x W x D)
Electrical equipment standard EN61010
Weight approx. 11 kg
FPM = Viton
®
PVDF = Polyvinylidenuoride
** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew point of 5 °C.
Komentarze do niniejszej Instrukcji