7
Gas sampling and gas conditioning technology 3-0.1-ME
6 INTRODUCTION
The VC-... pre-cooling units produced by M&C incorporates the “Jet-Stream’’ design of heat
exchanger. This design induces condensate formation and guarantees optimum dew point reduction to
ambient temperature.
7 APPLICATION
The M&C pre-cooling units VC-... are used in gas analysis to lower the dew point of humid gas, for
example:
• to relieve downstream main cooling system
• if process-bound water or steam irruption can become forward,
• if non-heated sample lines without adequate slope are mounted,
• for gas analysis system with electrochemical sensors.
8 TECHNICAL DATA
Pre-cooling unit version
VC-..
Jet-Stream heat exchanger
out of Duran glass EC-G
Autom. condensate
removal with SR25..
Gas flow rate
recommended, (other flow
rates possible)
Stagnant space per heat
exchanger
Terminals max. 2,5 mm
2
, PG11 cable gland
Electrical standard / Case
protection
Method of mounting /
Ready for working
Material of sample
contacting parts
Duran glass®,
PTFE, PVDF,
Novoprene
Duran glass®,
PTFE, PVDF,
Novoprene
Komentarze do niniejszej Instrukcji