9
Gas sampling and gas conditioning technology 3-4.1-ME
8 TECHNICAL DATA
range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
at constant conditions 0,25 °C
Gas flow rate per heat exchanger
Number of heat exchangers
1*, installation of max. 4 heat exchangers possible
Material of heat exchangers
Duran glass or PVDF or stainless steel 316
**+5 to +45°C
with option EC-F and EC-FD: +5°C to +50°C
with glass and PVDF: max. 3 bar
with stainless steel : 10 bar*
with option EC-FD : 1,2 bar
Dead volume per heat exchanger
Sample gas connection glass:
PVDF:
stainless steel:
For tube 6mm*,
option: 8mm, 10mm or hose connector
G 1/4“i,
option: hose connector
G 1/4“i*,
option: NPT or hose connector
Condensate connection glass:
PVDF:
stainless steel:
For tube 12mm*,
option: 8mm, 10mm or hose connector
G 3/8“i,
option: hose connector
G 3/8“i*,
option: NPT or hose connector
280VA, start up current at 230V = 7,9A
230V 50Hz +/- 10% *, or: 115V 60Hz +/- 10%
2 changeover contacts, max. 250V 2A AC/DC 500VA, 50W
19“ rack or wall mounting
EC : 84 TE x 7 HE x 450mm, 32kg
ECS: 84TE x 7 HE x 360mm, 31 kg
230V: R134A 100% CFC free
115V: R404A 100% CFC free
Electrical equipment standard
* standard
Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C
ambient temperature and an outlet dew point of 5 °C.